8163A/B, 8164A/B, 8166A/B User's Guide
E1x021: Invalid head caldata
The head calibration data was detected as invalid
x
refers to the module channel tested
Please return the unit to your local service center for repair
E1x031: 0.25V ref out of limits (a, b, c)
The 0.25V reference voltage measured through the 16bit ADC is out of the acceptable
limits
a
shows the ADC reading (0.25 = 3276)
b
shows the lower limit
c
shows the upper limit
x
refers to the module channel tested
Please return the unit to your local service center for repair
E10040: 10bit ADC test failed (a, b, c)
The 10bit ADC has returned a value other than 512
a
shows the ADC reading
b
shows the lower limit
c
shows the upper limit
Please return the unit to your local service center for repair