Keysight Technologies 8163A User Manual page 122

Lightwave multimeter, lightwave measurement system, lightwave multichannel system
Table of Contents

Advertisement

8163A/B, 8164A/B, 8166A/B User's Guide
You can generate input triggers in any of the following ways:
trigger the Input Trigger Connector on the rear panel of your
instrument,
set <Trigger Configuration> to <Loopback> so that an output trigger
automatically generates an input trigger, or
use the :TRIGger GPIB command, see your mainframe's Programming
Guide.
How to Use Output Triggering
You can configure 8163A/B Series Power Meter modules to output a
trigger when the instrument performs certain tasks.
To set your module's Output Trigger Configuration:
1
See
How to Set the Trigger Configuration
configure the trigger connectors.
2
Move to the Power Meter channel and press the [Menu] softkey.
3
Move to <Output Trigger Mode>, by using the cursor key, and press
Enter. You will see the screen in Figure 71.
Figure 71
4
Move to one of the following, by using the cursor key:
Output Trigger Mode
Power Measurement
on page 68 for how to
3
121

Hide quick links: