Keysight Technologies 8163A User Manual page 223

Lightwave multimeter, lightwave measurement system, lightwave multichannel system
Table of Contents

Advertisement

8
Setting Attenuation and/or Power Levels
222
4
Set the Attenuator module's
wavelength. Set the Attenuator module's
factor, to 0.000 dB.
5
Connect the source output to the Attenuator module input.
6
Use another patchcord to connect the Attenuator module output to
the reference powermeter input.
7
Note the insertion loss measurement displayed by the powermeter.
8
Select the details screen for the Attenuator.
9
Use the cursor key to select the
10 Use the cursor key to enter the insertion loss measured at
press [OK].
11 Disconnect the patchcord from the reference powermeter and connect
it to the DUT input.
Figure 136
81570A/71A/73A - Attenuation factor applied with no offset
Figure 136
on page -222 shows an attenuation factor
α
applied with no
.
Offset
λ
parameter to the appropriate
α
parameter, the attenuation
α
parameter, then press [Edit].
Offset
α
8163A/B, 8164A/B, 8166A/B User's Guide
Step
7, then
of 3.850 dB