E10060: Laser <a> control failed (b, c, d, e)
Checks the laser current source
a
laser number
b
measured current at low per and normal current source setting
c
measured current at low power and high current source setting
d
measured current at high power and normal current source setting
e
measured current at high power and high current source setting
Please return the unit to your local service center for repair
E10080: Laser <a> control for safety test failed
To test the current safety comparator the laser current needs to be set to two different
values
a
laser number
Please return the unit to your local service center for repair
E10081: Safety comparator <a> test failed (b, c)
Checks whether laser current safety comparator includes at the right conditions a
safety interrupt at the CPU
a
laser number
b
shows whether the interrupt flag was raised when the current was lower
than the limit
c
shows whether the interrupt flag was raised when the current was higher
than the limit
Please return the unit to your local service center for repair