error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
8163A/B, 8164A/B, 8166A/B User's Guide
E10052: Photodiode <a> temp out of limits (b)
The photodiode temperature is outside the allowed range
a
photodiode number
b
measured photodiode temperature
Verify whether the module is operated within the specified ambient temperature. If the
failure persists, please return the unit to your local service center for repair
E1x060: 0.25V offset failed (a, b)
Verifies that the offset voltage for the 16-bit ADC (250 mV) is correctly added to the input
signal
a
shows the ADC reading
b
shows the maximum difference to the exact ADC reading of 3277
x
refers to the internal PM channel tested
Please return the unit to your local service center for repair
E1x061: PGA DAC adj failed (a, b, c)
Tries to adjust the offset voltage of the PGA
a
shows the ADC Offset reading
b
shows the limit for an upper offset
c
shows the limit for a lower offset
x
refers to the internal PM channel tested
Please return the unit to your local service center for repair