16 A continuous sweep can end in two ways:
a when the sweep is completed, or
b when you press [Stop].
How to Perform a Manual Sweep
You can perform a manual sweep if you choose the <Manual> sweep mode
or if you press [Pause] during an automatic sweep. In a manual sweep you
choose when you want to perform each step of a stepped sweep. You can
choose to move forward or backward a wavelength or to end the sweep.
To perform a manual wavelength sweep over the range 1510 nm to
1570 nm, three times, sweeping two ways, in 1 nm steps:
1
Move to the Tunable Laser channel and press [Details].
2
Move to [Sweep Mode] and press Enter.
3
Move to the <Manual>, by using the cursor key, and press Enter.
step 4
4
Perform
5
Press the [Active] hardkey on the front panel of your Tunable Laser
module to enable the optical output.
6
Press [Run Swp] to start the sweep. The screen in
page -154 then appears.
7
You can:
press [Next] to move on to the next wavelength step,
press [Prev] to move on to the previous wavelength step, or
press [Stop] to end the sweep.
lambda logging can only be done in continuous sweep mode
the step size must be a multiple of the smallest possible step size
step 13
page 149
to
on
Tunable Lasers
to set the sweep parameters.