14
Error Messages
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
398
E1x062: PGA gain adj failed (a, b, c)
Verifies the gain settings of the PGA (x 1/10/100). A voltage of 25 mV is applied to the input
of the PGA and then amplified
a
if 1, the required division has failed
if 0, the calculated factors are correct
b
shows the measured values for an amplification by 10
c
shows the measured values for an amplification by 100
x
refers to the internal PM channel tested
Please return the unit to your local service center for repair
E1x063: Lambda Dac adj failed (a, b, c, d)
Verifies the gain of the responsiveness correction stage
a
if 1, the quotient d could not be calculated, otherwise 0
b
shows the difference between DAC settings 1024 and 2048
c
shows the difference between DAC settings 2048 and 4096
d
shows the quotient of b divided by c
x
refers to the internal PM channel tested
Please return the unit to your local service center for repair
E10070: Laser <a> control failed (b, c)
Checks the laser current source
a
laser number
b
low power setting
c
high power setting
Please return the unit to your local service center for repair