Keysight Technologies 8163A User Manual page 153

Lightwave multimeter, lightwave measurement system, lightwave multichannel system
Table of Contents

Advertisement

5
Tunable Lasers
152
6
Move to [
7
Enter 1560.000, press Enter.
8
Move to [Sweep Cycles] and press Enter.
9
Enter 0003 and press Enter.
10 Move to [Vsweep] and press Enter.
11 Move to 5 nm/s, by using the cursor key, and press Enter.
12 Move to [Step] and press Enter.
13 Enter 1.000 and press Enter.
For a <Continuous> sweep, [Step] sets the wavelength interval between
NOTE
output triggers, if you have set <Output Trigger Mode> to <Step
Finished>, see
14 Press the [Active] hardkey on the front panel of your Tunable Laser
module to enable the optical output.
15 Press [Run Swp] to start the sweep.
The following table explains all possible configuration problems:
NOTE
368,LambdaStop <=LambdaStart
369,sweepTime < min
370,sweepTime > max
371,triggerFreq > max
372,step < min
373,triggerNum > max
374,LambdaLogging = On AND
Modulation = On AND
ModulationSource! =
CoherenceControl
375,LambdaLogging = On AND
TriggerOut! = StepFinished
λ
Stop] and press Enter.
How to Use Output Triggering
start wavelength must be smaller than stop wavelength
the total time of the sweep is too small
the total time of the sweep is too large
the trigger frequency (calculated from sweep speed divided by sweep
step) is too large
step size too small
the number of triggers exceeds the allowed limit
The only allowed modulation source with the lambda logging function is
coherence control.
lambda logging only works "Step Finished" output trigger configuration
on page 165.
8163A/B, 8164A/B, 8166A/B User's Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a8163b8164b8166a8166b81636b

Table of Contents