Power Measurement
112
Figure 64
Out of Range - Power Greater Than Upper Power Limit
If the measured power is more than 40 dBm less than the range setting, it
is impossible for power to be displayed. The power value, -1.--- dBm, as
shown in
Figure 65
on page -113, is shown. This means that the measured
power is greater than the resolution at this [Range] value. You must
increase the [Range] value.
See
Table 11
on page -114 for more details.