14
Error Messages
action
error message
description
test value
action
406
b
shows the time when the temperature regulation has reached the default
temperature. Waits maximum 30 seconds
x
refers to the module channel tested
Please return the unit to your local service center for repair
E1x081: TEC current to temperature mismatch
If the TEC temperature difference does not correspond to the measured Peltier current
this test will fail
x
refers to the module channel tested
Verify whether the module is operated within the specified ambient temperature. If the
failure persists, please return the unit to your local service center for repair.
8163A/B, 8164A/B, 8166A/B User's Guide