Error Messages
error message
description
test value
action
error message
description
test value
action
402
E1x050: TEC NTC temp ou of limits (a, b, c, d)
The sensor temperature is out of the acceptable limits
a
shows the sensor NTC temperature
b
shows the sensor NTC temperature ADC reading
c
shows the lower limit
d
shows the upper limit
x
refers to the module channel tested
Verify whether the module is operated within the specified ambient temperature. If the
failure persists, please return the unit to your local service center for repair.
E1x051: TEC current out of limits (a, b, c)
Verifies whether the TEC current is within a given range
a
shows the ADC reading of the TEC current
b
shows the lower limit
c
shows the upper limit
x
refers to the module channel tested
Please return the unit to your local service center for repair
8163A/B, 8164A/B, 8166A/B User's Guide