Figure 176
Setting Analysis Parameters
22 To edit the power units used for the analysis, move to [W], press Enter,
move to <dBm>, <W>, or <dB>, and press Enter. See
Power Units ?
23 To edit the reference mode used for the analysis, move to <Ref Mode>,
press Enter, move to one of the following referencing modes:
<Value>, you set a reference value using <Ref>, see
<First Sample>, the first sample of the stability function is selected
as the reference value,
<Channel 1>, the value that Channel 1 measures is updated as the
reference value for each set of power measurements (thus, each
data value for channel 1 will be 0 dB), and
press Enter.
These settings have an effect when you choose <dB> as the <Pwr unit>.
NOTE
See step 7 or step 22.
The Keysight 8163B Stability Analysis Screenn - First Screen
on page 101 for an explanation of power units.
What are the
step 24
8163A/B, 8164A/B, 8166A/B User's Guide
,