8163A/B, 8164A/B, 8166A/B User's Guide
Initialization Error (TASK_VERIFY, id1, id2)
A task could not be started
id1
task name
id2
maximum available memory
Reboot the instrument. If the failure persists, please return the unit to your local service center
for repair
If one of the selftests fails, please try to reboot the instrument. If the
selftest continues to fail, please return the mainframe to your local service
center for repair.
Initialization Error (Tcfs Amplifier Calibration failed, error value)
The calibration of the tuning control feedback system amplifier has failed
0
(not significant)
Please return the unit to your local service center for repair
Initialization Error (TcfsAdpEEPROMInit, error value)
Error reading tuning control feedback system adapter board EEPROM
1
(not significant)
Please return the unit to your local service center for repair
Initialization Error (TcfsEEPROMInit, error value)
Error reading tuning control feedback system EEPROM
1
(not significant)
Please return the unit to your local service center for repair