E10051: Laser <a> temp out of limits (b)
The laser temperature is outside the allowed range
a
laser number
b
measured laser temperature
Verify whether the module is operated within the specified ambient temperature. If the
failure persists, please return the unit to your local service center for repair
E10060: Laser <a> control failed (b, c, d, e)
Checks the laser current source
a
laser number
b
measured current at low power and normal current source setting
c
measured current at low power and high current source setting
d
measured current at high power and normal current source setting
e
measured current at high power and high current source setting
Please return the unit to your local service center for repair
E10070: Peltier <a> temp diff amp failed
Checks whether the comparator output changes from low to high values while the
temperature DAC is set to two different temperatures
a
laser number
Please return the unit to your local service center for repair