Keysight Technologies 8163A User Manual page 404

Lightwave multimeter, lightwave measurement system, lightwave multichannel system
Table of Contents

Advertisement

error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
8163A/B, 8164A/B, 8166A/B User's Guide
E1x052: TEC comparator test failed (a, b, c, d)
Checks the TEC regulation voltage comparator
a
shows the 10bit ADC reading for the high output voltage
b
shows the 10bit ADC reading for the low output voltage
c
shows the lower limit
d
shows the upper limit
x
refers to the module channel tested
Please return the unit to your local service center for repair
E1x053: TEC default temp invalid (a)
Verifies whether the default temperature is within a given range
a
shows the default temperature
x
refers to the module channel tested
Please return the unit to your local service center for repair
E1x054: TEC control failed (a, b, c)
When the TEC is enabled the temperature must move towards the default temperature,
coming from a previous temperature before the TEC was enabled. The moving direction
must be correct and the overshoot must be less than a certain limit
a
shows the temperature before the TEC was enabled
b
shows the actual temperature while TEC is enabled
c
shows the target temperature, where b should move to
x
refers to the module channel tested
Please return the unit to your local service center for repair
Error Messages
14
403

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a8163b8164b8166a8166b81636b

Table of Contents