error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
8163A/B, 8164A/B, 8166A/B User's Guide
E1x052: TEC comparator test failed (a, b, c, d)
Checks the TEC regulation voltage comparator
a
shows the 10bit ADC reading for the high output voltage
b
shows the 10bit ADC reading for the low output voltage
c
shows the lower limit
d
shows the upper limit
x
refers to the module channel tested
Please return the unit to your local service center for repair
E1x053: TEC default temp invalid (a)
Verifies whether the default temperature is within a given range
a
shows the default temperature
x
refers to the module channel tested
Please return the unit to your local service center for repair
E1x054: TEC control failed (a, b, c)
When the TEC is enabled the temperature must move towards the default temperature,
coming from a previous temperature before the TEC was enabled. The moving direction
must be correct and the overshoot must be less than a certain limit
a
shows the temperature before the TEC was enabled
b
shows the actual temperature while TEC is enabled
c
shows the target temperature, where b should move to
x
refers to the module channel tested
Please return the unit to your local service center for repair