error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
8163A/B, 8164A/B, 8166A/B User's Guide
Initialization Error (Calibration Data Invalid, error value)
The calibration data was tagged as invalid before it was loaded on to the module
0
not significant
Please return the unit to your local service center for repair
Initialization Error (CalOptZeroMeas, error value)
Error in optical zero measurement
1
not significant
Please return the unit to your local service center for repair
Initialization Error (DspMainCalcKFactor failed, error value)
DSP K factor calibration failed
-1
no TCFS (according to caldata)
1
error when calibrating TCFS
2
DSP sweep failed
4
laser safety
Please return the unit to your local service center for repair
Initialization Error (DspMainWriteCaldata(), error value)
Error while writing cal data to DSP
5
tried to access invalid array position
6
wrong dimension
7
unknown type found in variable array
8
wrong parameter when trying to access one of the DSPs cal
arrays
Please return the unit to your local service center for repair