Keysight Technologies 8163A User Manual page 385

Lightwave multimeter, lightwave measurement system, lightwave multichannel system
Table of Contents

Advertisement

14
Error Messages
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
384
Initialization Error (Firmware CRC wrong: crc1, stored: crc2!)
The checksum stored in the firmware does not match the calculated checksum
crc1
calculated checksum
crc2
checksum read from caldata
Reload the firmware of the module. If the failure persists, please return the unit to your local
service center for repair
Initialization Error (HwAccessInit, error value)
Error setting DAC
1
(not significant)
Please return the unit to your local service center for repair
Initialization Error (InitDacs, error value)
Error setting DAC
1
(not significant)
Please return the unit to your local service center for repair
Initialization Error (InstSettingInit, error value)
Error creating semaphore
2
(not significant)
Please return the unit to your local service center for repair
Initialization Error (LaserEEPROMInit, error value)
Error reading laser EEPROM
1
(not significant)
Please return the unit to your local service center for repair