Usb modular 5.5 digits digital multimeter (45 pages)
Summary of Contents for Keysight Technologies U1281A
Page 1
Keysight U1281A/U1282A Handheld Digital Multimeter Programming Guide...
Page 2
WITHOUT NOTICE, IN FUTURE EDITIONS. without prior agreement and written applicable in any technical data. FURTHER, TO THE MAXIMUM EXTENT consent from Keysight Technologies as PERMITTED BY APPLICABLE LAW, governed by United States and Safety Notices KEYSIGHT DISCLAIMS ALL WARRANTIES, international copyright laws.
Handheld Digital Multimeter Programming Guide Introduction Configuring the IR interface Introduction to Remote Commands Syntax conventions Querying parameter settings Returned result Prompts This chapter describes the parameters that configure the U1281A/U1282A, and helps you determine the settings to achieve optimum performance.
Introduction Configuring the IR interface The U1281A/U1282A may be operated via an IR interface. The remote control operation will enable you to manually operate the multimeter via a terminal, or to automatically execute a host computer program. The default parameters for the U1281A/U1282A are: –...
TRIGger. Only the short form—which is TRIG—will be recognized by the instrument. Other forms, such as TRIGG, TRIGge, or even TRIGger, will generate errors. An example of the command syntax format is illustrated below: -> TRIG:SOUR <BUS|REF|IMM> Keysight U1281A/U1282A Programming Guide...
-> CONF? Responding Message Returned result After the U1281A/U1282A executes a query command, the return of the result will be in the following format: <- <Result> + <CR> <LF> On the multimeter warning the returned prompts will be in the following format except Xon and Xoff: <- <Prompts>...
Page 9
Rotary position: CAP/TEMP measurement ﹡ 6 Rotary position: uA/mA measurement ﹡ 7 Rotary position: A measurement ﹡ 8 ﹡ 9 Rotary position: SQU output ASC(&H11) Xon: The meter is available ASC(&H13) Xoff: The meter is busy Keysight U1281A/U1282A Programming Guide...
Page 10
Introduction THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight U1281A/U1282A Programming Guide...
Refer to “U1281A/U1282A Configuration Overview” on page 12. For more details, refer to the U1281A/U1282A User’s Guide. Due to the unique Export Log feature of the U1281A/U1282A, you will need to handle the NOTE exporting of data when using any of the following functions: –...
440mA FUSED Connect the test leads to the appropriate input terminals according to the selected measurement function. When the above configuration has been made, you can proceed to use the remote commands in the following sections. Keysight U1281A/U1282A Programming Guide...
*IDN? Syntax *IDN? Description The *IDN? query allows the U1281A/U1282A to identify itself by returning the model number, serial number, and the firmware version. The string returned is, for example: Keysight Technologies,<model number>,<serial number>,VXX.xx where: – <model number> identifies which U1281A/U1282A model is being used.
Syntax *RST Description The *RST (reset) command places the U1281A/U1282A in the power-on-reset state without shutting down the line power and returns the current rotary knob position (refer to “Prompts” on page for a legend on rotary knob positions). The interface parameters will return to their default values (refer to “Configuring the IR interface”...
Remote Commands *CLS Syntax *CLS Description The *CLS (clear status) command clears the event registers in all register groups. This command also clears the Error queue. Example The following command clears the event -> *CLS register bits. Keysight U1281A/U1282A Programming Guide...
This command places the trigger hold system in the idle state. The display will show a dash (–) until the INIT command is triggered. Example Places the trigger hold mode in the idle state -> ABOR Remarks This command is only available in trigger hold mode. Keysight U1281A/U1282A Programming Guide...
– MIN returns the minimum recorded value – AVER returns the average recorded value – PRES returns the present recorded value – COUN returns the count value of the average mode Example Returns the maximum recorded value -> CALC:AVER:MAX? <- +2.63782000E+00 Keysight U1281A/U1282A Programming Guide...
Query CALC:PEAK:MAX|MIN? Description This query returns the value of the peak-hold function. – MAX returns the maximum recorded value – MIN returns the minimum recorded value Example Returns the maximum peak-hold value -> CALC:PEAK:MAX? <- +2.83752000E+00 Keysight U1281A/U1282A Programming Guide...
The query is returned in the following format string: <“Function”, “Range”, “Resolution”>. When the @2 parameter is omitted, the query returns the function of the primary display. Example Returns the function of the primary display -> CONF? <- VOLT:AC +6.00000000E+01,+1.00000000E-03 Keysight U1281A/U1282A Programming Guide...
This query returns the value of the primary or secondary display. When the @2 parameter is omitted, the query returns the value of the primary display. The @3 parameter corresponds to the environment temperature. Example Returns the value of the primary display. -> FETC? <- +1.23475000E+00 Keysight U1281A/U1282A Programming Guide...
Removes the trigger hold mode from the idle -> INIT state and re-triggers the system. Returns the value of the primary display. -> FETC? <- +1.23266000E+00 Remarks This command is only available in trigger hold mode. Keysight U1281A/U1282A Programming Guide...
Example Enters the trigger hold mode. -> TRIG:SOUR:BUS Returns a value after the re-trigger cycle when in the trigger hold mode, or the display -> READ? value when not in the trigger hold mode. <- -1.20000000E-01 Keysight U1281A/U1282A Programming Guide...
Remote Commands STAT? Syntax STAT? Description This query returns the status of the U1281A/U1282A in the following format string: <ABCDEFGHIJKLMNOPQRSTU> Each of the letters represented in the format string are explained below: Letter Parameter Description 0 : Off Average function...
Page 24
Battery level 1 : Low 0 : Five decimal places Resolution 1 : Four decimal places 0 : Off AC Low Pass Filter status 1 : On 0 : Off DC Filter status 1 : On Keysight U1281A/U1282A Programming Guide...
Page 25
Remote Commands Example Returns the status of the U1281A/U1282A -> STAT? <- 000000000910L00200000 Keysight U1281A/U1282A Programming Guide...
Remote Commands SYST:BATT? Syntax SYST:BATT? Description This query returns the residual capacity of the battery in the U1281A/U1282A. Example Returns the residual capacity of the battery -> SYST:BATT? currently in the U1281A/U1282A. <- 100% Keysight U1281A/U1282A Programming Guide...
Enters the auto hold mode. -> TRIG:SOUR:REF Returns the type of the trigger source. -> TRIG:SOUR? <- REF Remarks The average and peak functions are disabled when you enter either trigger hold mode or auto hold mode. Keysight U1281A/U1282A Programming Guide...
Sets a count of 50 to the auto hold mode. -> TRIG:REF:COUN 50 Returns the auto hold mode count. -> TRIG:REF:COUN? <- +5.00000000E+01 Remarks The count will be stored in the non-volatile memory, but the “0” parameter will not. Keysight U1281A/U1282A Programming Guide...
Need help?
Do you have a question about the U1281A and is the answer not in the manual?
Questions and answers