8163A/B, 8164A/B, 8166A/B User's Guide
Figure 87
How to Execute a Continuous Sweep
Continuous sweep mode is not available for all tunable laser models.
NOTE
Only the 81602A, 81606A, 81607A, 81608A and 81960A allow choosing
a repeat mode for a continuous sweep. To manually select two way
continuous sweeps for these models, first select Stepped Sweep mode,
then select two way Repeat mode, and then return to Continuous Sweep
mode.
To execute a continuous wavelength sweep over the range 1520 nm to
1560 nm, three times, at a speed of 5 nm/s:
1
Move to the Tunable Laser channel and press [Details].
2
Move to [Sweep Mode] and press Enter.
3
Move to the <Continuous> sweep mode and press E
4
Move to [
5
Enter 1520.000 and press Enter.
Pausing a Stepped Sweep
λ
Start] and press Enter.
Tunable Lasers
5
.
NTER
151