Keysight Technologies 8163A User Manual page 405

Lightwave multimeter, lightwave measurement system, lightwave multichannel system
Table of Contents

Advertisement

14
Error Messages
error message
description
test value
action
error message
description
test value
action
404
E1x060: 0.25V offset failed (a, b)
Verifies that the offset voltage for the 16-bit ADC (250 mv) is correctly added to the
input signal
a
shows the ADC reading
b
shows the maximum difference to the exact ADC reading of 3277
x
refers to the module channel tested
Please return the unit to your local service center for repair
E1x061: PGA DAC adj failed (a, b, c)
Tries to adjust the offset voltage of the PGA
a
shows the ADC offset reading
b
shows the limit for an upper offset
c
shows the limit for a lower offset
x
refers to the module channel tested
Please return the unit to your local service center for repair
8163A/B, 8164A/B, 8166A/B User's Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a8163b8164b8166a8166b81636b

Table of Contents