These settings have an effect when you choose <dB> as the <Pwr unit>.
NOTE
See step 7 or step 22.
13 If you have chosen <Value> as the <Ref Mode>, move to <Ref>, press
Enter, enter a value for the reference, and press Enter.
Setting Stability Parameter
14 Move to [Total Time], press Enter. [Total Time] is a value of the form
xx:yy:zz, where xx is a value in hours (maximum 23 hours), yy is a value
in minutes (maximum 59 minutes), and zz is a value in seconds
(maximum 59 seconds).
Enter a two-digit time value in hours, press the right cursor key, enter a
two-digit time value in minutes, press the right cursor key, enter a
two-digit time value in seconds, and press Enter.
15 Move to [AvgTime], press Enter, move to an averaging time period, and
press Enter.
16 Move to [Max], press Enter, enter an integer value between 1 and
4000, and press Enter.
Running a Stability Application
17 You can run the stability application by pressing [Measure]. The
Stability Measurement screen, as shown in
appears. A progress bar shows the proportion of the stability
application that has been completed.
Figure 173
on page -281,