8
Setting Attenuation and/or Power Levels
228
The difference between the power level measured by your reference
powermeter and the power value measured by your Attenuator P
stored by the Attenuator module as P
P
.
actual
4
If the reference powermeter is hosted by another mainframe:
Manually enter the value of P
P
= P
- P
Off
actual
ref
where P
is the power level measured by your reference
ref
powermeter.
5
Disconnect the input to the reference powermeter, then connect it to
the DUT input.
6
Edit the P
parameter to control the absolute power applied to the
SET
DUT. The attenuation,
value.
Figure 138
81576A/77A Setting P
, then applied to P
Offset
Off
α
, displayed is updated to reflect the new P
after calibration using [PM->Off]
SET
8163A/B, 8164A/B, 8166A/B User's Guide
is
actual
and
SET
SET