Keysight Technologies 8163A User Manual page 391

Lightwave multimeter, lightwave measurement system, lightwave multichannel system
Table of Contents

Advertisement

14
Error Messages
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
390
TCFS Temp. Selftest Failed (error value)
Verifies if the peltier temperature is within target with a given uncertainty
15010
TCFS peltier temperature is not in target range
Please return the unit to your local service center for repair
TCFS Resistor Selftest Failed (error value)
This test measures the R
of the photodiodes at channel A, B and M
d
15021
Photodiode R
in channel A is below limit
d
15022
Photodiode R
in channel B is below limit
d
15023
Photodiode R
in channel C is below limit
d
Please return the unit to your local service center for repair
TCFS Attenuator Selftest Failed (error value)
Verifies whether an attenuator position is repeatable.
15030
Attenuator could not be positioned
Please return the unit to your local service center for repair
Main DC Selftest Failed (error value)
Tests all fixed DC voltages available at the mainboard mux
23011
+15V Measurement DC voltage not in limits
23012
-15V Measurement DC voltage not in limits
23013
+5V Reference DC voltage not in limits
23014
-5V Reference DC voltage not in limits
23015
negative supply voltage of laser current source not in limits
23016
positive pezo supply voltage not in limits
Please return the unit to your local service center for repair
8163A/B, 8164A/B, 8166A/B User's Guide