Keysight Technologies 8163A User Manual page 386

Lightwave multimeter, lightwave measurement system, lightwave multichannel system
Table of Contents

Advertisement

error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
8163A/B, 8164A/B, 8166A/B User's Guide
Initialization Error (MainEEPROMInit, error value)
Error reading main EEPROM
1
(not significant)
Please return the unit to your local service center for repair
Initialization Error (MotorSelfCali failed, error value)
Motor self calibration failed
-1
(not significant)
Please return the unit to your local service center for repair
Initialization Error (MotorZero, error value)
DSP motor zero not finished in time
1
(not significant)
Please return the unit to your local service center for repair
Initialization Error (OptZero, error value)
Optical zero disabled in caldata
-1
(not significant)
Please return the unit to your local service center for repair
Initialization Error (OptZero Opc, error value)
Timeout while waiting for laser diode (16) and thermopositioner (64) to settle while preparing
optical zero measurement
16
waiting for laser diode
64
waiting for thermopositioner
Reboot the instrument. If the failure persists, please return the unit to your local service center
for repair
Error Messages
14
385