3
Power Measurement
114
Figure 66
Upper Power Limit and Resolution
Table 11
on page -114 shows the upper power limit and measurement
resolution at various power ranges. As can be seen the resolution
decreases as the chosen [Range] decreases. The resolution is always
40 dBm less than the chosen [Range] value. The Upper Power Limit is
always 3 dBm greater than the chosen [Range] value.
Table 11
[R
]
Upper Power Limit
ANGE
40 dBm
19.999 W
30 dBm
1.9999 W
20 dBm
199.99 mW
10 dBm
19.999 mW
Range Value Menu
Upper Power Limits and Resolution for Various Power Ranges
43.000 dBm
33.000 dBm
23.000 dBm
13.000 dBm
Resolution
0.001 W
-0.000 dBm
0.1 mW
-10.000 dBm
0.01 mW
-20.000 dBm
0.001 mW
-30.000 dBm
8163A/B, 8164A/B, 8166A/B User's Guide