Keysight Technologies 8163A User Manual page 418

Lightwave multimeter, lightwave measurement system, lightwave multichannel system
Table of Contents

Advertisement

action
error message
description
test value
action
8163A/B, 8164A/B, 8166A/B User's Guide
b
shows the measured values for an amplification by 10
c
shows the measured values for an amplification by 100
Please return the unit to your local service center for repair
E10063: Lambda Dac adj failed (a, b, c, d)
Verifies the gain of the responsiveness correction stage
a
if 1, the quotient d could not be calculated, otherwise 0
b
shows the difference between DAC settings 1024 and 2048
c
shows the difference between DAC settings 2048 and 4096
d
shows the quotient of b divided by c
Please return the unit to your local service center for repair
The following error codes are attenuator module specific:
Error Creating message queue!
Error Init MCB!
Error spawning tLowlvlMsgHandler!
Error spawning tKeyPressed!
Error spawning tTempCtrl!
Error spawning tPowCtrl!
Error checking firmware CRC!
Error Messages
14
417

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a8163b8164b8166a8166b81636b

Table of Contents