Keysight Technologies 8163A User Manual page 165

Lightwave multimeter, lightwave measurement system, lightwave multichannel system
Table of Contents

Advertisement

5
Tunable Lasers
164
Figure 93
Input Trigger Mode
4
Move to one of the following, by using the cursor key:
<Ignore>, input triggers are ignored.
<Next Step>, an input trigger will cause the next step of a stepped
sweep to be performed.
<Run Sweep>, an input trigger will start a single sweep cycle.
5
Press Enter.
You can generate input triggers in any of the following ways:
applying a trigger to the Input Trigger Connector on the rear panel
of your instrument,
setting <Trigger Configuration> to <Loopback> so that an output
trigger automatically generates an input trigger, or
using the :TRIGger GPIB command, see your mainframe's
Programming Guide.
8163A/B, 8164A/B, 8166A/B User's Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a8163b8164b8166a8166b81636b

Table of Contents