Keysight Technologies 8163A User Manual page 417

Lightwave multimeter, lightwave measurement system, lightwave multichannel system
Table of Contents

Advertisement

14
Error Messages
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
error message
description
test value
416
E10031: 0.25 ref out of limits (a, b, c)
The 0.25V reference voltage measured through the 16bit ADC is out of the acceptable
limits
a
shows the ADC reading (0.25V = 3276)
b
shows the lower limit
c
shows the upper limit
Please return the unit to your local service center for repair
E1060: 0.25V offset failed (a, b)
Verifies that the offset voltage for the 16-bit ADC (250 mV) is correctly added to the input
signal
a
shows the ADC reading
b
shows the maximum difference to the exact ADC reading of 3277
Please return the unit to your local service center for repair
E10061: PGA DAC adj failed (a, b, c)
Tries to adjust the offset voltage of the PGA
a
shows the ADC Offset reading
b
shows the limit for an upper offset
c
shows the limit for a lower offset
Please return the unit to your local service center for repair
E10062: PGA gain adj failed (a, b, c)
Verifies the gain settings of the PGA (x 1/10/100). A voltage of 25 mV is applied to the input
of the PGA and then amplified
a
if 1, the required division has failed
if 0, the calculated factors are correct
8163A/B, 8164A/B, 8166A/B User's Guide