Keysight Technologies 8163A User Manual page 233

Lightwave multimeter, lightwave measurement system, lightwave multichannel system
Table of Contents

Advertisement

8
Setting Attenuation and/or Power Levels
232
If your Attenuator module is not hosted by the same mainframe as your
NOTE
reference powermeter you can, at step 7, enter the appropriate Offset
manually.
During a test that utilizes the
When the Offset (
the operating wavelength
λ
the
Offset Table entries, as shown in
Offset Value
extrapolation:
same value for
all l < l
Figure 140
If the operating wavelength matches a wavelength stored in the
Offset Table, the exact offset value is used. The Attenuator status
display includes 'Exa'.
If the operating wavelength is between two wavelengths stored in the
Offset Table, the offset applied is calculated by linear interpolation. The
Attenuator status display includes 'Int'.
If the operating wavelength is greater than, or less than, the range of
wavelengths stored in the
wavelength is extrapolated. The Attenuator status display includes
'Ext'.
λ
) feature is on the exact offset applied depends on how
λ
you set for the Attenuator module relates to
linear interpolation
1
l
1
Extrapolation and Interpolation of Offset Values
λ
λ
Offset Table
Figure 140
on page -232.
Offset Table, the offset value for the closest
8163A/B, 8164A/B, 8166A/B User's Guide
extrapolation:
same value for
all l > l
n
l
l
n
λ
λ