Hssl Failure - Receiver Hardware Error/Bl; Monitor Power-On Self-Test Error/Bl - Abbott CELL-DYN 3200 System Operator's Manual

Table of Contents

Advertisement

Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
10-52
Table 10.2 General Fault Conditions (Continued)
HSSL failure – receiver hardware error/BL
The HSSL receiver on the Data Module malfunctioned.
– – – - Probable Cause – – –
1. A hardware failure on the HSSL
card has occurred.

Monitor power-on self-test error/BL

NOTE: This message pertains to a software execution monitor on the Analyzer,
not the video display monitor.
(e.g., test pattern not successfully read from memory)
– – – - Probable Cause – – –
1. A failure in the Analyzer hardware
has occurred.
1. Initialize the Analyzer by pressing
the [INITIALIZATION] key on the
second DIAGNOSTICS MENU
screen. If the message appears
again, power OFF the Analyzer.
Then reseat both ends of HSSL
cable and make sure the
connections are secure. Power
Analyzer ON.
NOTE: For the location of the
NOTE: If you wish to use the
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
A failure occurred during self-test.
1. Reboot the System.
NOTE: If you wish to use the
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Section 10
– – – - Corrective Action –
HSSL connectors on the
rear panel of the Data
Module, refer to Section
1: Use or Function;
Subsection: Data Module
Components.
Sample Loader, reset the
racks before initializing or
rebooting the System.
– – – - Corrective Action –
Sample Loader, reset the
racks before rebooting the
System.
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001