Section 6
Entering New Calibration Factors — Closed Mode
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001
1. In the CALIBRATION menu, press [ENTER FACTOR].
2. Use the arrow keys to select the first factor in the Closed Sampler column to
be changed. Type in the New Calibration Factor calculated in Calculating
New Closed Mode Calibration Factors above. Press the Enter key to save the
factor and advance the cursor. If necessary, use the arrow keys to select the
next factor to be changed.
NOTE: Prior to exiting the screen, the operator may press the [RESTORE
FACTORS] soft key to recall factors, stored on the Hard Disk,
corresponding to the current mode — Open or Closed. [RESET ALL
TO 1.000] is used to reset all factors displayed on the screen to
1.000.
3. When all new factors have been entered, press [RETURN] to display the
CALIBRATION LOG screen with the new calibration factors.
NOTE: If no changes are made in the ENTER CALIBRATION FACTOR screen,
pressing [RETURN] will bypass the CALIBRATION LOG screen and
display the main CALIBRATION menu.
4. Complete the Calibration Log (adding comments) as required. If a printout of
the log is desired, press [PRINT LOG].
5. Press [RETURN] twice to return to the main CALIBRATION menu.
6. Press [PRINT] to obtain a copy of the new Closed Sampler Calibration Factors.
7. Press [MAIN] to return to the MAIN MENU.
Calibration Procedures
6-47