Abbott CELL-DYN 3200 System Operator's Manual page 152

Table of Contents

Advertisement

Performance Characteristics and Specifications
Sample Loader Specifications
Sample Loader Operational Specifications
4-20
Operational specifications for the Sample Loader are given in Table 4.20.
Table 4.20
Sample Loader Operational Specifications
Feature
Rack Capacity
Tube Capacity
Sample Throughput
(1) Patient
(2) FWBC, RRBC, QC
Run Cycle Time
(1) Patient
(2) FWBC, RRBC, QC
Number of Mixing Stations
Degree of Mixing Rotation
Number of Mixing Rotations
Maximum Tube Size
Bar Code Reader
Bar Code Labels
For information on tube types and size, refer to Tube Types in Section 13: Sample
Loader; Subsection: Physical and Performance Specifications.
For information on bar codes types and operation, refer to Appendix A: Bar
Codes.
For general information on Sample Loader operation, refer to Section 13: Sample
Loader; Subsection: Sample Loader Description.
Specification
5 racks on the load side;
5 racks on the unload side
50 (10 tubes per rack)
(1) 58 samples/hour
(2) 34 samples/hour
(1) 62 seconds
(2) 105 seconds
2: Mixing Stations #1 and #2
0 – 135 degrees
15 times at Mixing Station #1
15 times at Mixing Station #2
16mm (DIA) X 80mm (High)
LED type
Code 39, I2of5, Codabar, Code 128
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001