Download Flow Sequence Timeout/Bl; Command Transmission Timeout - Abbott CELL-DYN 3200 System Operator's Manual

Table of Contents

Advertisement

Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
10-56
Table 10.2 General Fault Conditions (Continued)

Download flow sequence timeout/BL

The allotted time for flow sequence downloading was exceeded.
– – – - Probable Cause – – –
1. The flow sequence did not compile,
possibly due to interruption by a
fatal fault, an improperly seated
Tower Cover or door, or a fluidics
problem.
The Data Module did not receive acknowledgment of a command sent to the
Analyzer within the allotted time.
– – – - Probable Cause – – –
1. A failure in the Analyzer hardware,
possibly due to an unplugged or
damaged HSSL cable, has occurred.
2. A software error has occurred.
1. Reboot the System.
NOTE: If you wish to use the
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.

Command transmission timeout

1. Check the HSSL cable connections.
Then, initialize the Analyzer by
pressing the [INITIALIZATION] key
on the second DIAGNOSTICS MENU
screen. If the message appears
again, reboot the System.
NOTE: For the location of the
NOTE: If you wish to use the
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Section 10
– – – - Corrective Action –
Sample Loader, reset the
racks before rebooting the
System.
– – – - Corrective Action –
HSSL ports, refer to
Section 1: Use or
Function; Subsection:
Data Module
Components.
Sample Loader, reset the
racks before initializing or
rebooting the System.
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001