Abbott CELL-DYN 3200 System Operator's Manual page 320

Table of Contents

Advertisement

Calibration Procedures
Auto-Cal Method
6-22
Figure 6.7
START AUTO CAL Key Displayed
A message is displayed on the bulletin line alerting the operator to start running
samples.
As samples are processed, the results of each run are displayed as shown in
Figure 6.8. The following soft keys are displayed:
ACCEPT
DELETE A RUN
QUIT
PRINT
RETURN
(Label appears after the 5th run)
(Label appears after the 1st run)
(Label appears after the 1st run)
CELL-DYN
Section 6
®
3200 System Operator's Manual
9140181H—October 2001