Section 10
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001
Conditions (Continued)
Table 10.1
Status
Loader Warning 145: unload area full/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The unload area contains 5 racks. NOTE: The Sample Loader halts.
No corrective action is necessary. When you wish to run additional samples in
the Sample Loader, remove the racks from the unload area and press the [START
LOADER] key.
Loader Warning 146: unload area nearly full/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The unload area contains 4 racks.
No corrective action is necessary. If a rack is being processed and there are no
other faults, the Sample Loader will continue to operate.
Loader Status 158: load zone empty/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The load zone empty sensor is not sensing the presence of a rack in the load zone.
NOTE: The Sample Loader halts.
When you wish to run samples in the Sample Loader, load tubes in racks and
press the [START LOADER] key.
Troubleshooting and Diagnostics
10-37