Section 6
(1)
New Open
Mode Factor
WOC
NOC
RBC
HGB
MCV
PLT
1. In column 1, enter the new factor calculated in column 4 of the previous worksheet.
2. In column 2, enter the calibration factor that existed prior to running the current calibration procedure.
3. Subtract the current factor in column 2 from the new factor in column 1 and divide the result by the current
factor. Enter this result in column 3.
4. Multiply the result in column 3 by 100 and enter the result in column 5.
(1)
Factor%
Diff
WOC
NOC
RBC
HGB
MCV
PLT
1. In column 1, enter the new Factor % Diff from column 5 of the previous worksheet (disregard the sign).
2. If the new Factor % Diff exceeds the limit in column 4, DO NOT CALIBRATE. Call the Abbott
Hematology Customer Support Center for assistance.
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001
Worksheet 2
Open Mode Factor % Difference
(Calculate All Factors To Three Decimal Places)
(2)
Current Open
Mode Factor
Worksheet 3
Open Mode Calibration Range Criteria
(2)
Lower Limit
Cal Not Required
<1.5%
<1.5%
<1.0%
<1.0%
<1.0%
<3.0%
(3)
/
New Open
Mode Factor
/
/
/
/
/
/
(3)
Calibration Range
Cal Required
>1.5% but <10%
>1.5% but <10%
>1.0% but <10%
>1.0% but <10%
>1.0% but <10%
>3.0% but <15%
Calibration Procedures
(4)
(5)
Factor
x 100 =
% Diff
x 100 =
x 100 =
x 100 =
x 100 =
x 100 =
x 100 =
(4)
Upper Limit
Do Not Cal
Cal? Y/N
>10%
>10%
>10%
>10%
>10%
>15%
(5)
6-53