Service and Maintenance
Special Protocols
Reagent Reservoir
9-4
CLEAN/RES SHEAR VAL/
CLEAN SHEAR VAL/
RESTORE SHEAR VAL
DIS/ ENAB ANALYZER/
DISABLE ANALYZER/
ENABLE ANALYZER
MORE
MAIN
A brief description of the function of each soft key follows. Instructions for the
detailed use of each key are given in the appropriate maintenance procedure.
Figure 9.2
Reagent Reservoir Keys
NOTE: This procedure is performed only in the Open Mode. If necessary, press
[CHANGE SAMPLER] in the RUN screen to select Open Mode.
The [REAGENT RESERVOIR] key is used to drain the reagent reservoirs located on
the Flow Panel of the Analyzer. When the [REAGENT RESERVOIR] key is pressed,
the following soft key labels are displayed as shown in Figure 9.2:
EMPTY DIL/SHEATH
FILL DIL/SHEATH
EMPTY HGB LYSE
FILL HGB LYSE
EMPTY WBC LYSE
FILL WBC LYSE
RETURN
.H\ ODEHO DOWHUQDWHV EHWZHHQ VHOHFWLRQV
(.H\ ODEHO DOWHUQDWHV EHWZHHQ VHOHFWLRQV
(The key alternates between the selections.)
(The key alternates between the selections.)
(The key alternates between the selections.)
CELL-DYN
Section 9
®
3200 System Operator's Manual
9140181H—October 2001