Abbott CELL-DYN 3200 System Operator's Manual page 627

Table of Contents

Advertisement

Section 14
Retic Diagnostics Menu
FAULT
REPORT
RETIC
CNT GRAPH
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001
Move Specimen
The [MOVE SPECIMEN] key is used to move a QC result located in one file
(indicated by the cursor position) to another QC file. When the [MOVE SPECIMEN]
key is pressed, the RETIC QC LOG screen is displayed, allowing the desired file to
be selected. When the [MOVE TO FILE] key is pressed, the result is moved to the
indicated file.
Procedure: Move Specimen
1. From the RETIC QC LOG screen, use the arrow keys on the keyboard to move
the cursor to the file containing the result to be moved.
2. Press the [VIEW QC LOG] key.
3. Use the arrow keys on the keyboard to position the cursor at the result that is
to be moved.
4. Press the [MOVE SPECIMEN] key to display the VIEW RETIC QC LOG menu.
5. Use the arrow keys on the keyboard to move the cursor to the file in which
the results are to be placed.
6. Press the [MOVE TO FILE] key to move the results to the designated file.
NOTE: The results are moved to the end of the list of data that is currently
in the file. The placement of results in the list of data will determine
their location on the Levey-Jennings graph.
7. The VIEW RETIC QC LOG screen for the original file is displayed showing that
results have been moved.
Print QC Log
The [PRINT QC LOG] key is used to print the entire QC Log.
Return
The [RETURN] key is used to return to the RETIC QC LOG screen.
EXECUTION
RETIC
TIMES
CNT RATE
PRINT
RETURN
RETIC
CNT RATE
CD3200 RETIC MENU
DIAGNOSTICS MENU
RETIC
CLEAR
RAW DATA
FAULTS
PRINT
RETURN
Reticulocyte Package
RETIC
MAIN
14-39