Troubleshooting and Diagnostics
Diagnostics Menu
Execution Times
CNT Rate Summary
10-6
Operator correctable faults (for example, waste full, diluent empty) can be cleared
by pressing the [CLEAR FAULT] key after taking corrective action. After corrective
action for a fatal fault, the system must be reinitialized by pressing the
[INITIALIZATION] soft key in the second DIAGNOSTICS MENU screen (refer to
Figure 10.9).
This key is for service use only.
Figure 10.5 Count Rate Summary Screen
When the [CNT RATE SUMMARY] key is pressed, the following soft key labels (see
Figure 10.5) are displayed:
WOC CNT RATE/WOC CNT GRAPH/WOC CNT DATA *
RBC/PLT CNT RATE/RBC/PLT CNT GRAPH/RBC/PLT CNT DATA *
NOC CNT RATE/NOC CNT GRAPH/NOC CNT DATA *
PRINT
RETURN
* Key labels alternate between the selections.
NOTE: Once a [CNT RATE] key is pressed, that key toggles between [CNT DATA]
and [CNT GRAPH] until the operator exits the screen. The keys then revert
to [CNT RATE].