Flow Sequence Time Out <X,X>/Bl; Command Sent At Incorrect Time <X>/Bl - Abbott CELL-DYN 3200 System Operator's Manual

Table of Contents

Advertisement

Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
10-46
Table 10.2 General Fault Conditions (Continued)
Flow sequence time out <x,x>/BL
NOTE: Characters in the brackets identify the problem flow sequence number
and name.
– – – - Probable Cause – – – –
1. An internal Analyzer flow sequence
was not completed in the allotted
time.
Command sent at incorrect time <x>/BL
– – – - Probable Cause – – – –
1. The Analyzer received a command
from the System Software at the
incorrect time and could not process
it.
– – – - Corrective Action – – –
1. Record the characters in the
brackets.
2. Initialize the Analyzer by pressing
the [INITIALIZATION] key on the
second DIAGNOSTICS MENU
screen. Continue processing
samples.
3. If the problem recurs, contact your
Abbott Hematology Customer
Support Center.
– – – - Corrective Action – – –
1. Record the characters in the
brackets.
2. Initialize the Analyzer by pressing
the [INITIALIZATION] key on the
second DIAGNOSTICS MENU
screen. Continue processing
samples.
3. If the problem recurs, contact your
Abbott Hematology Customer
Support Center.
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001
Section 10