Service and Maintenance
Nonscheduled Maintenance Procedures
9-58
b. The skirt on the SL model has two removable sections adjacent to the
Flow Panel. Remove the section on the left side of the Flow Panel by
lifting it up and off its holding pins.
7. Carefully remove the tubing from all the Normally Closed Valves on the
Flow Panel. Do not detach the tubing. If necessary, refer to Figure 9.4 for the
location of these valves.
8. Carefully remove the tubing from behind the Sample Transfer Peristaltic
Pump. Do not detach the tubing. If necessary, refer to Figure 9.6.
9. Remove all three Reagent Tubing lines, Waste Tubing line, and the Waste
Sensor line from their fittings on the rear of the Analyzer. The waste line
should be emptied and rinsed with disinfectant.
10. Place each length of tubing in a separate protective plastic bag and close the
bag. (Keep the Waste line and Waste Sensor line together.)
NOTE: Keep all tubing lines clean to prevent contamination.
11. Place the plastic bags containing the reagent and waste tubing in the
Accessory Kit.
12. Wipe the instrument with the disinfectant prepared earlier.
NOTE: If the instrument will be shipped, the following steps should also be
done.
13. Remove and clean the Shear Valve as directed in Shear Valve Cleaning under
Nonscheduled Maintenance Procedures.
14. After the Shear Valve has been cleaned and dried, wrap the ceramic center
section carefully for protection and place it in the Accessory Kit.
15. Obtain the Shear Valve Dummy Center section from the Accessory Kit.
Reassemble the Shear Valve on the instrument using the dummy center
section.
16. Remove the Analyzer power cord from its connector on the rear of the
Analyzer and from the outlet receptacle. Place the cord in the Accessory Kit.
17. Remove the Display Monitor power cord from its outlet receptacle and
secure it to the monitor.
18. Detach the Display Monitor Video Cable and the Membrane Keypad
Interface Cable from their respective connectors on the Data Module and
secure the cables to the monitor.
19. Detach the Data Module Interface Cable from the back of the Analyzer.
20. Replace covers:
a. On the CS model, reattach the Front Skirt using the four Phillips-head
screws.
b. On the SL model, align the removable left side skirt panel over its
holding pins and slide down.
c. Close Left and Right Front Covers as follows. Lift door fully upward
from bottom area near door hinge bracket. While still holding door up,
close door far enough to allow door to drop down into its normal
position. Gently lower door and press into place so magnetic latch is
engaged.
d. Reattach the Tower Cover and press into place (Tower Door on the CS
model must be open).
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001
Section 9