Reticulocyte Package
Retic Menu Options
QC FILE SETUP
14-20
Procedure: Range Entry
1. Select a file from the RETIC QC LOG screen by using the arrow keys on the
keyboard to move the cursor to the desired file.
2. Press the [RETIC QC LIMITS] key and if necessary press the [RANGE ENTRY]
key to display the RETIC QC RANGE ENTRY screen for the selected file.
3. Use the arrow keys on the keyboard to move the cursor to the desired entry
field.
4. Type the desired numbers and press the Enter key on the keyboard to save
each entry.
5. If desired, press the [PRINT] key to obtain a printout of the entered values.
6. Press the [RETURN] key to save the entries and return to the RETIC QC LOG
screen.
NOTE: When entries are saved, the software checks to see if any entries
would result in the upper limit being less than the lower limit. If this
situation occurs, the limits are automatically reversed. The bulletin
line displays the following message:
LIMITS WERE EXCHANGED TO MAKE UPPER > LOWER
Procedure: Means/Limits Entry
1. Select a file from the RETIC QC LOG screen by using the arrow keys on the
keyboard to move the cursor to the desired file.
2. Press the [RETIC QC LIMITS] key and if necessary press the [MEANS/LIMITS]
key to display the RETIC QC MEANS/LIMITS entry screen for the selected file.
3. Use the arrow keys on the keyboard to move the cursor to the desired entry
field.
4. Type the desired mean and limit values and press the Enter key on the
keyboard to save each entry.
5. If desired, press the [PRINT] key to obtain a printout of the entered values.
6. Press the [RETURN] key to save the entries and return to the RETIC QC LOG
screen.
The following soft key labels are displayed on the QC FILE SET UP screen:
LOT NUMBER or
REPLICATE ID
TOGGLE ON/OFF
PRINT
RETURN
The [TOGGLE ON/OFF] key is used to configure the QC file. The file can be used for
commercial controls. The lot number and expiration date may be entered into the
appropriate areas on the QC FILE SET UP screen (see the Figure 14.9) and saved by
pressing the Enter key on the keyboard. The Westgard
toggled ON or OFF using the [TOGGLE ON/OFF] key.
(This key label alternates between
these two selections.)
Rule selections can be
®