Abbott CELL-DYN 3200 System Operator's Manual page 282

Table of Contents

Advertisement

Operating Instructions
Using The Data Log
Reject From X-B
5-130
If the cursor is positioned at a sample identified with a "r", "w", or "b" preceding
the sequence number (indicating that the results are included in the XB analysis),
the [REJECT FROM X-B] key label is displayed. (See Figure 5.47.)
Figure 5.47 Data Log Screen Showing Reject From X-B Key
When the [REJECT FROM X-B] key is pressed, the sample is marked with an "R"
following the Specimen ID, the results are excluded from the X-B analysis (the
"R", "w", or "b" is deleted), and the key label changes to [ACCEPT INTO X-B]. (See
Figure 5.48.)
If the [ACCEPT INTO X-B] key is pressed, the "R" is deleted, the "r", "w", or "b" is
displayed and results are now included in the X-B analysis.
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001
Section 5