Operating Instructions
Using The Data Log
5-140
6. Repeat steps 3–5 until all selections have been made.
7. If desired, press the Print Screen key on the keyboard to obtain a printout of
the selected groups.
8. Press [RETURN] to return to the DATA LOG screen.
9. The Data Log is displayed configured with the selected parameters.
Standard Groups
The Data Log display may also be customized with pre-determined groups
(Standard Groups) of parameters using the [STANDARD GROUPS] key. Figure 5.54
shows the WBC Group placed in GROUP 1, the RBC Group placed in GROUP 2,
the PLT Group placed in GROUP 3 and the Diff Group placed in GROUP 4.
To Customize the Data Log Display for Standard Groups
1. From the main DATA LOG screen, press [CUSTOMIZE DATA LOG] to display the
CUSTOMIZE DISPLAY FOR DATA LOG screen.
2. Press [STANDARD GROUPS] to display the CUSTOMIZE DISPLAY FOR DATA
LOG screen and key labels for Standard Groups.
3. Use the Arrow keys on the keyboard to move the cursor to the desired group
(1–4) location.
NOTE: This number indicates the order in which the group of parameters
will be displayed (Group 1 on the first screen, Group 2 on the
second, etc.).
4. Press the soft key corresponding to the desired parameter group. This group
is displayed in the position indicated by the cursor.
5. Repeat steps 3 and 4 until all desired groups have been selected.
6. If desired, press the Print Screen key on the keyboard to obtain a printout of
the configuration.
7. Press [RETURN] to return to the DATA LOG screen.
8. The Data Log is displayed configured with the standard groups of
parameters.
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001
Section 5