Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
10-58
Table 10.2 General Fault Conditions (Continued)
The buffer for Analyzer-to-Data Module communication overflowed.
– – – - Probable Cause – – –
1. The Analyzer message rate was too
high for the Data Module to keep up
with.

Analyzer is not communicating its mode/BL

The Analyzer is not communicating whether it is an SL or CS model.
– – – - Probable Cause – – –
1. An Analyzer software setup error
has occurred.

Tower/Loader Fault 3: transmission failure/BL

A transmission acknowledgment failure occurred between the Analyzer and the
module that controls the tower and loader.
– – – - Probable Cause – – –
1. A failure of the electronics or
communication hardware has
occurred.

Overflow in the HSSL queue/BL

1. Reboot the System.
NOTE: If you wish to use the
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
1. Reboot the System.
NOTE: If you wish to use the
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
1. Reboot the System.
NOTE: If you wish to use the
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Section 10
– – – - Corrective Action –
Sample Loader, reset the
racks before rebooting the
System.
– – – - Corrective Action –
Sample Loader, reset the
racks before rebooting the
System.
– – – - Corrective Action –
Sample Loader, reset the
racks before rebooting the
System.
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001