Abbott CELL-DYN 3200 System Operator's Manual page 67

Table of Contents

Advertisement

Section 1
Top Panel
Flow Cell Assembly
Laser Optics Bench Assembly
1 Orthogonal (90° and
90°D) Scatter Light
Detectors
2 Laser Tube
3 Forward Angle (0° and
10°) Scatter Light
Detectors
4 Optical Flow Cell
5 Laser
Figure 1.10 Optical Bench
CELL-DYN
3200 System Operator's Manual
®
9140181K—July 2002
The Flow Cell Assembly contains the fluidics and hardware needed to
hydrodynamically focus the RBC/PLT and WBC sample streams in the path of the
laser beam for analysis. The primary components of this assembly are:
• Sample Feed Nozzle — a specially designed tube used to deliver the diluted
sample into the sheath stream
• Sample Flow Cell — an optically clear quartz chamber with a central square
opening of a specific size, which flares out into a cone at the bottom of the
flow cell
The Laser Optics Bench Assembly (located behind the Flow Panel) contains the
Helium-Neon laser, the flow cell, and the optics and detectors required for
enumeration and differentiation of white blood cells, red blood cells, and platelets
(refer to Figure 1.10).
CAUTION –
CLASS 3B LASER LIGHT WHEN OPEN.
AVOID EXPOSURE TO BEAM.
PN 9230701
5
1
4
Use or Function
2
3
1-19