Reticulocyte Package
Troubleshooting
Operational Messages and Data Flagging
Dispersional Data Alerts
Instrument Alert Messages
Table 14.2
Instrument Alert Messages
Alert
Flow Error
NOTE: Alert occurs when
the average count rate
rapidly increases during the
Reticulocyte count cycle.
Too Few Events
NOTE: Alert occurs when
fewer than 3000 events are
counted during the
Reticulocyte count cycle.
>>>> (Chevrons)
14-68
The result of each run (patient, control, or background) is reviewed within the
appropriate limits as entered by the operator or taken from the instrument's preset
limits. If a result for a parameter exceeds these limits, they are flagged on the screen
and on the report. Dispersional Data Alerts are displayed or printed as follows:
Data Station Screen Display:
• Result(s) below lower limits shown in yellow
• Result(s) above upper limits shown in purple
Linearity Exceeded:
Reticulocyte QC Log:
Reticulocyte Data Log:
Graphics Report:
Instrument Alert Messages are displayed when the instrument detects an
inappropriate condition during specimen processing. When necessary, data is
suppressed. When these messages occur, follow the instructions given, and take the
appropriate corrective action. When the problem is corrected, repeat the specimen.
Instrument Alert Messages with Suppressed Reticulocyte Results
Suppression of Reticulocyte results occurs when the sample run data acquisition
process exceeds normal parameters. When the Reticulocyte results are suppressed,
one of the following three alerts will be displayed in the lower left-hand quadrant
of the Display Monitor screen in the Reticulocyte Package and on the graphics
printout under the heading ALERTS.
Probable Cause
· Air bubble
· Hardware malfunction
· Inadequate whole
blood sample mixing
- Improper pipetting
- Blood not stained
· Cold Agglutinin
· The Reticulocyte
percentage exceeds the
reportable linear range.
Result displayed as >>>>
Result(s) outside limits underlined when printed
Result(s) outside limits underlined when printed
Result(s) outside limits underlined
Corrective Action
1. Run a background count to cycle air through
the system.
2. Rerun the Reticulocyte specimen.
3. If alert still occurs, refer to Section 10:
Troubleshooting and Diagnostics;
Subsection: Troubleshooting Flow Errors.
1. Prepare another dilution verifying proper
specimen preparation as discussed in
Routine Operation, Reticulocyte
Specimens, Running Specimens, Patient,
Specimen Preparation within this chapter.
2. Verify Reticulocyte results by an alternate
method.
1. Verify Reticulocyte results by an alternate
method.
CELL-DYN
Section 14
3200 System Operator's Manual
®
9140181K—July 2002