Loader Fault 154: Tube Position Bar Code Read Failure/Bl - Abbott CELL-DYN 3200 System Operator's Manual

Table of Contents

Advertisement

Section 10
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001
Table 10.2 General Fault Conditions (Continued)

Loader Fault 154: tube position bar code read failure/BL

A tube position bar code label (the second through tenth labels on the rack) could
not be read.
– – – - Probable Cause – – –
1. The window on the Bar Code
Reader is dirty.
2. The tube position bar code label is
scuffed or dirty.
Troubleshooting and Diagnostics
– – – - Corrective Action –
1. Remove the Loader Cover. Using a
DYN-A-WIPE pad or lint-free
wipe moistened with isopropyl
alcohol, lens cleaner or deionized
water, clean the window on the Bar
Code Reader. Dry the window on
the Bar Code Reader thoroughly.
Then, reinstall the Loader Cover,
reset the racks, and reset the Sample
Loader by pressing the following
keys in order:
[CLEAR FAULT],
[START LOADER],
[RESET LOADER].
NOTE: For the location of the Bar
Code Reader, refer to
Section 13: Sample
Loader; Subsection:
Sample Loader
Description.
2. Clean or replace the tube position
bar code label. Then, reset the racks
and reset the Sample Loader.
If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
10-81