Service and Maintenance
Nonscheduled Maintenance Procedures
1 Tube Fitting
2 Luer Lock
3 Holding Block
4 Diluent/Sheath
Syringe
5 Double Collar
Figure 9.17 Diluent/Sheath Syringe Replacement
9-44
1
2
3
4
5
9. Insert the double collar on the plunger into the slot on the mounting bracket
and line up the circular flange on the barrel with the slot on the holding block.
10. Insert one of the vertical edges on the barrel into a side groove on the holding
block and carefully twist the barrel until the other side snaps into place. Make
sure the syringe is firmly in place.
11. Place the tube fitting into the Luer Lock and turn the lock counterclockwise
until the fitting is finger tight. Be careful not to overtighten the fitting or
crimp the associated tubing.
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001
Section 9