Section 9
CELL-DYN
®
3200 System Operator's Manual
9140181H—October 2001
Figure 9.19 Open Sample Probe Replacement
8. Hold the probe steady and reinsert the tubing over the top. Use the pliers to
carefully work the tubing down the probe. Be sure the tubing is firmly seated
on the probe.
NOTE: Wetting the top portion of the probe will allow the tubing to slide
more easily.
9. Press [ENABLE ANALYZER] followed by [RETURN], [MAIN], and [RUN].
10. Press the Touch Plate to run a background count. Observe the action of the
probe assembly to ensure there are no leaks and that it is operating smoothly.
11. Close the Right Front Cover. Reattach the Tower Cover and press into place
(Tower Door on the CS model must be open).
12. Run at least three background counts to rinse the system. Check that the
background counts are acceptable before running controls or patient samples.
If the counts are unacceptable, troubleshoot accordingly (refer to Section 10:
Troubleshooting and Diagnostics; Subsection: Troubleshooting Guide).
13. Record this maintenance in your maintenance log.